4-1
Chapter 4
Installation
4.1 Environmental Issues
Ideally, temperature calibration equipment should be used in a calibration
laboratory or other facility specifically designed for this purpose. Environmental
requirements include:
Stable temperature and relative humidity < 80 %
Clean, draft-free area
Low noise level: low radio frequency, magnetic or electrical interference
Low vibration levels
4.2 Mounting
Most often temperature standards, primary and secondary, are used to calibrate
other temperature-sensitive equipment. The PRT must be mounted carefully to
avoid any damage to the sheath or sensor. If the fluid bath used does not have a
lid designed for PRT insertion, clamps should be used to ensure the handle and
cable are not immersed. Do not screw the clamps too tight. Over tightening will
damage the sheath. If metal comparison blocks are used in the bath, maintain a
close fit between the thermometer sheath and the well in the comparison block.
However, allow for the thermal expansion of the thermometer sheath when
determining block well tolerances.