2271A
Operators Manual
34
Connect a UUT
Connect the UUT to the top test port on the Product using the method described
below:
W Caution
To prevent Product damage or damage to a gauge:
• DO NOT use PTFE tape on these connections. This prevents
a correct seal. The Gauge Adapter sealing system can be
hand-tight sealed up to 20 MPa (3000 psi). Wrenches or
similar tools are not required. Over tightening can cause
damage to threads or sealing faces.
• Before connection, ensure that there is an O-ring fitted to
the test port.
• Check that the sealing face of the device to be fitted is clean
and undamaged, as scratches or dents can form leak-paths.
Note
The thread on the test port, and the lower part of the gauge adapters
is LEFT-HANDED. The following procedure details the correct
method for mounting devices using these adapters:
1. Screw the appropriate gauge adapter fully on to the UUT, see Figure 6.
hwr029
Figure 6. Screw on Gauge Adapter
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com