7-1
Chapter 7
Storage & Transportation
Instrument
1. Shut-off any external supplies, and depressurize system by slowly opening both Inlet
and Exhaust valves.
2. Disconnect external supplies.
3. Remove screw-in handle from handpump arm (if fitted) and store in tool roll.
4. Refit instrument lid, ensuring that the hinges are correctly engaged, and secure with
toggle clips at the sides.
Weights
1. Starting with the largest increments first, stack all appropriate weights in the wooden
weight case.
2. Pass the threaded rod of the weight clamp assembly down through the center of the
weight stack, and locate in the base of the weight box.
3. Secure the weights by screwing the clamp assembly in clockwise, ensuring that the
stepped rim of the clamping disc locates correctly in the center of the weight stack.
4. Close lid and secure with catches at front.
W Warning
The weight set is heavy (can be up to 80 lbs/36 kg per individual
box) so care must be taken when moving it. Both handles must
be used when lifting the set for stability, and we recommend
that it be carried between two people.