8808A
Users Manual
1-8
General Specifications
Voltage
100V Setting ...................................................... 90 V to 110 V
120V Setting ..................................................... 108 V to 132 V
220V Setting ..................................................... 198 V to 242 V
240V Setting ..................................................... 216 V to 264 V
Frequency.......................................................... 47 Hz to 440 Hz
Power Consumption ......................................... 15 VA peak (10 W average)
Dimensions
Height................................................................. 88 mm (3.46 in)
Width.................................................................. 217 mm (8.56 in)
Depth ................................................................. 297 mm (11.7 in)
Weight ................................................................ 2.1 kg (4.6 Ib)
Display
Vacuum Fluorescent Display, segment
Environment
Temperature
Operating ........................................................ 0 °C to 50 °C
Storage ........................................................... -40 °C to 70 °C
Warm Up......................................................... ½ hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating ........................................................ <90 % (0 °C to 28 °C)
<75 % (28 °C to 40 °C)
<45 % (40 °C to 50 °C)
Storage .......................................................... -40 °C to 70 °C <95 %
Altitude
Operating ....................................................... 2,000 Meters
Storage............................................................... 12,000 Meters
Vibration ............................................................ Complies with MIL-PRF-28800F Class 3
Safety
Complies with IEC 61010-1:2001, ANSI/ISA 61010-1 (S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2 No.
61010.1:2004, CAT I 1000V/CAT II 600 V
EMC
Designed to comply with IEC 61326-1:1997+A1:1998+A2:2000
Triggering
Trigger Delay .................................................... 400 ms
External Trigger Delay ..................................... <2 ms
External Trigger Jitter ..................................... <1 ms
Trigger Input ..................................................... TTL Levels
Trigger Output ................................................... 5 V max
Math Functions
Min/max, relative, hold, compare and dB functions
Electrical
Input Protection ............................................... 1000 V all ranges