FLUKE SK-700HPM-FS Pneumatic Test Pump Kit

Instructions Manual - Page 2

For SK-700HPM-FS.

PDF File Manual, 6 pages, Read Online | Download pdf file

SK-700HPM-FS photo
Loading ...
Box Contents
The Product ships with
700HPP High Pressure Pneumatic Test Pump
700HPM High Pressure Calibration Manifold with
hose (700HPPK only)
Canvas carrying bag
High Pressure Pneumatic Quick Service Kit
Dry pac refill
Silico ne grease
High Pressure hose
Instructions
Fittings (as listed in Table 2)
Table 2. Pump Fittings
Configuration
Ordered
M16- ¼
NPT
(Male)
M16-
NPT
(Male)
M16-¼
BSP
(Male)
M16-
BSP
(Male)
M16-
M14
(Male)
¼ NPT
(F)-
M20x
1.5
(Male)
700HPP-
NPT
1 1 - - - -
700HPPK-
NPT
700HPP-
BSP
- - 1 1 - -
700HPPK-
BSP
700HPP-
MET
1 - - - 1 1
700HPPK-
MET
Safe Operation
To verify safe operation of the Product after maintenance
or repair:
1. Wear eye protection and securely attach a pressure
gauge to the hose or accessory fitting of the
calibration manifold (700HPPK only). Use any
approved (properly-rated) adapter as required.
2. Close the upper vent valve (700HPPK) and slowly
pressurize the pump/gauge system to 21 MPa
(3000 psi, 210 bar).
3. Maintain pressure for 60 seconds.
4. Use the upper vent valve on the calibration manifold
(700HPPK only) to slowly release the pressure.
Pressure can be vented using the vent valve at the
base of the pump when using the 700HPP.
Features
The features and components of the Product are shown in
Figure 1 and Table 3.
A
B
3
4
5
6
7
8
9
10
11
12
13
A
B
14
15
1
2
16
23
21
22
17
18
19
20
hza01.eps
Figure 1. Features and Components
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...