Milliamp Process Clamp Meter
Introduction
XW Warning
To avoid electric shock or personal injury, do not perform the
calibration verification tests or calibration procedures
described in this manual unless you are qualified to do so.
The information provided in this manual is for the use of
qualified personnel only.
This manual provides the complete verification and adjustment procedure for the
771 Milliamp Process Clamp Meter (referred to in this manual as the Meter).
The Meter allows closed-case calibration using reference sources. It measures the
reference signals, calculates the correction factors, and stores them in memory. The
instrument should be calibrated after repair, or if it fails a performance test.
The 771 Calibration Manual provides the following information:
• Precautions and safety information
• Specifications
• Basic maintenance
• Calibration verification procedure
• Replaceable parts and accessories
For complete operating instructions, refer to the 771 Instruction Sheet.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com