87V MAX
Users Manual
40
Detailed Specifications
For all detailed specifications:
Accuracy is specified for 2 years after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 95 %.
Accuracy specifications take the form of ±([% of Reading] + [Number of least-significant digits]). For the 4 ½-digit mode, multiply the number
of least-significant digits (counts) by 10.
AC Voltage
AC conversions are ac-coupled and valid from 3 % to 100 % of range.
Range Resolution
Accuracy
45 – 65 Hz 15 – 200 Hz 200 – 440 Hz 440 Hz – 1 kHz 1 – 5 kHz 5 – 20 kHz
600.0 mV 0.1 mV
±(0.7 % + 4)
±(1.0 % + 4)
[1]
±(2 % + 4)
±(2 % + 20)
[2]
6.000 V 0.001 V
60.00 V 0.01 V
±(0.7 % + 2)
±(2 % + 4)
[3]
Unspecified
600.0 V 0.1 V Unspecified
1000 V 1 V Unspecified Unspecified
Low-Pass Filter
±(1.0 % + 4)
[1]
+1.0 % + 4
-6.0 % - 4
[4]
Unspecified Unspecified Unspecified
[1] Below 30 Hz, use smoothing function. Below 20 Hz add 0.6 %.
[2] Below 10 % of range, add 12 counts.
[3] Frequency range: 1 to 2.5 kHz
[4] Specification increases from -1 % to -6 % at 440 Hz when filter is used.