SVA1000X User Manual 74
2.6.5.3 Ref Level
Set the reference level to indicate the minimum value that the current grid can display. This
value is also displayed in the status bar on the left side of the screen.
2.6.5.4 Scale
Set the vertical scale of each grid to adjust the range of amplitude that can currently be
displayed. This value is also displayed in the status bar on the left side of the screen.
2.6.5.5 Calibration
Set calibration related items. This function is only available when the measurement item is S11.
There are three calibration states, factory calibration, user calibration, and no calibration. The
user calibration status is displayed in green font at the top left corner of the screen. Enter the
sub-menu; you can make the appropriate choice.
1. 1-Port Cal
Calibrating with the specified mechanical calibration requires three loads: open, short, and
match. Calibration data is saved as user calibration data.
2. Response (Through)
When operating, connect the port 1 and port 2 of the analyzer with an optional Through
Adapter. The normalization operation moves the measurement reference plane to both ends of
the Through Adapter. This function is only available when the measurement item is S21. Enter
the normalized submenu to make the appropriate selection.
3. Ecal (not available)
Use the optional SIGLENT electronic calibration kit for calibration. Calibration data is saved as
user calibration data.
4. Clear
Clear user calibration data.
5. Cal kit
Specifies the calibration kit used for mechanical calibration.
F503ME:Type-N 50Ω 3 GHz Cal Kit (Siglent)
85032F:Type-N 50Ω 9 GHz Cal Kit (KeySight)
6. Modify Cal kit (not available)
Use user defined calibration kit.