3700 SERIES USER MANUAL
3700 Series Programmable DC Electronic Load
VERSION 2.0 2006.03
8
Constant resistance setting procedure:
Procedure Operation details LCD display
Step 1 Press “R-set” button Input Password
Step 2
Enter the password (if the keyboard isn’t locked, please
go to step4)
Input Password
****
Step 3
Press the “OK ” button, the original set value will be
displayed on the LCD. (It will return to step 2 if your
password is wrong)
Set Resis = 200.0Ω
New=
Step 4
Enter the new resistance value with the number keypad
or rotate the dial to adjust the value
Set Resis = 200.0Ω
New= 50.00
Step 5
Press “R-set” button for confirmation
Set Resis = 50.00Ω
New= 50.00
Step 6
Press ESC to go back to the main menu
You can exit the resistance set operation at any point by pressing the ESC key
2.2.4 Store a program
The 3700 Series electronic load provides a program mode for dynamic testing. The user can program the 3700
series electronic load and define the load values (current, power, resistance) and duration of every step. For more
details about the programming operation, please refer to 2.3.7.
User’s program may be stored in the 3700’s nonvolatile memory for later use. Up to 9 programs can be stored
in the 3700. To store a program, please follow this operation procedure:
Procedure Operation details LCD display
Step1 Press “ Store ” button SAVE 1
Step 2
Enter the program number (1 to 9) by using the number
key or the rotary knob
SAVE *
Step 3
Press the “ OK ” button for confirmation. It will return
to Step 2 if the number is out of the range.
You can exit the store operation at any point by pressing the ESC key