C.Technique Parameter:
D.Operating Steps:
1. Connect the circulating pump before startup (if there is no water inlet and outlet, this step
can be omitted):
1.1 for the connection of circulating pump during internal circulation, connect the water
outlet and water inlet directly with both ends of the hose supplied with the goods.
1.2 for the connection of circulating pump during external circulation, connect the water
outlet to the inlet of container / equipment outside the tank with hose, and connect the water
inlet to the outlet of container / equipment outside the tank.
2. Add the corresponding liquid medium into the tank. The liquid level of the liquid medium
shall not be 20mm lower than the worktable to avoid damage caused by exposed dry burning
of the heater.
2.1 When the working temperature is lower than 8 , the liquid medium is generally
industrial alcohol or antifreeze.
2.2 when the working temperature is 8 - 75 , the liquid medium is generally purified
water.
2.3 when the working temperature is above 75 , dimethyl silicone oil with viscosity of
50cs is generally selected as the liquid medium.
3Plug in the power, turn on the "main power" switch, and then turn on the "power" button on
the operation panel. (Note: "main power" switch is on the back of the instrument).
4Perform the instrument operation as follows:
4.1Display Description:
Model
Temperature
range
Temperature
fluctuation
±℃
Slot
opening
mm
Groove
depth
(mm)
Power
KW
DC-0506
-5-100
0.05
180*160
150
0.8