PSO2020
Users Manual 12
1. Attach the tip of probe and ground nip to the Connector of Probe. A square wave
will be displayed within a several seconds. (Approximately 1 kHz, 2V, peak- to-
peak).
1.7 Self Calibration
The self calibration routine lets you optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run the
routine if the ambient temperature changes by 5 or more. For accurate calibration,
power on the oscilloscope and wait twenty minutes to ensure it is warmed up. To
compensate the signal path, disconnect any probes or cables from the input connectors.
Then, access the “Menu -> Utility -> Calibration” option and follow the directions on the
screen. The self calibration routine takes about several minutes.
1.8 Accessories
All the accessories listed below are standard accessories for the oscilloscope:
A PC software CD of the oscilloscope