Loading ...
Loading ...
Loading ...
SIGLENT
32 SPD3000X QuickStart
8. OUTPut
Command format
OUTPut {CH1|CH2|CH3},{ON|OFF}
Description
Turn on/off the specified channel output.
Example
OUTPut CH1,ON
Command format
OUTPut:TRACK {0|1|2}
Description
Select operation mode. Parameters {0|1|2} mean
independent, series and parallel respectively.
Example
OUTPut: TRACK 0
Command format
OUTPut:WAVE {CH1|CH2},{ON|OFF}
Description
Turn on/off the Waveform Display function of specified
channel.
Example
OUTPut:WAVE CH1,ON
9. TIMEr
Command format
TIMEr:SET
{CH1|CH2},{1|2|3|4|5},<voltage>,<current>,<time>
Description
Set timing parameters of specified channel, including
group{1|2|3|4|5}, voltage, current, time
Example
TIMEr:SET CH1,2,3,0.5,2
Command format
TIMEr:SET? {CH1|CH2},{1|2|3|4|5};
Description
Query the voltage/current/time parameters of specified
group of specified channel.
Loading ...
Loading ...
Loading ...