ADC 432 Adtemp Mini Non-Contact Infrared Thermometer

User Manual - Page 2

For 432.

PDF File Manual, 2 pages, Download pdf file

432 photo
Loading ...
How to Use
NOTE: Charge thermometer for two hours before first use.
1. Clean the unit before using.
2. Make sure the site is clean and dry prior to taking measurement.
3. Press and release the multifunction button to turn on. You will
hear a short beep and the unit will perform a self check.
A second beep will indicate that the unit is ready to use.
4. Aim probe at the measuring site at a distance of 1-2 cm.
5. Press the multifunction button to take measurement.
6. When measurement is complete,
the unit will beep and the
color-coded temperature will appear
(see Figure 1).
7. The unit will power off automatically
in about 10 seconds.
Memory Function
The unit retains the last nine readings.
To recall previous readings:
1. While the unit is off, press and hold the multifunction button for about three seconds,
until the unit enters memory mode.
2. Release the button. The reading number will appear first, then the corresponding
measurement value. These numbers will toggle three times.
3. Press and release the multifunction button again to recall each earlier reading.
4. The unit will switch to measurement mode about three seconds after the last button push.
To delete all readings in memory:
1. While in memory mode, press and hold the multifunction button for about three seconds,
until the “del” code appears on the display.
2. Release the button.
3. Press and release the multifunction button once more to confirm deletion. All readings
will be deleted and the unit will power off.
Charging the Battery
Charge the unit when the low-battery indication appears.
Do not dispose of the battery in a fire. The battery may explode.
1. When the low-battery indication shows, charge it via micro USB cable.
2. During charging, the low-battery icon lights with yellow; once
charging is completed, the low-battery icon lights with green.
3. Unplug when charging is complete.
Cleaning and Disinfecting
The thermometer can be cleaned, and an intermediate level of disinfection can be achieved
using the following method.
CAUTION:
Never use abrasive cleaning agents, thinners, or benzene for cleaning and never
immerse the instrument in water or other cleaning liquids.
CAUTION:
Never insert a sharp object into the sensor lens area or any other open surface on
the thermometer.
CAUTION:
Do not use unapproved cleaning or disinfection agents. Use of these agents may
cause damage to components.
CAUTION:
Do not use chemicals other than isopropyl or ethyl alcohol on the sensor lens.
Cleaning the sensor lens
Slightly moisten a cotton swab or cloth with isopropyl or ethyl alcohol and gently wipe the
surface of the sensor lens using a side-to-side (not circular) motion. Avoid touching the
sensor lens except when cleaning is required.
Cleaning the thermometer
Use a pre-moistened wipe or cloth slightly moistened with an approved cleaner to gently
clean the thermometer.
Disinfecting the thermometer
Use a pre-moistened wipe or cloth slightly moistened with an approved disinfectant to disinfect
the thermometer. After cleaning or disinfecting, wait at least 10 minutes before taking another
measurement. Follow the manufacturer’s recommendations.
Calibration Testing
The thermometer is calibrated at the time of manufacture. If the thermometer is operating
according to these instructions, periodic readjustment is not required. ADC recommends
checking calibration every two years or whenever clinical accuracy of the thermometer is
in question. ADC can perform these tests for a nominal fee.
Care and Storage
Do not drop or crush; this device is not shock proof.
Do not dismantle or modify device.
Do not store the device in an extreme environment with direct sunlight or high/low
temperatures, or in a location with high humidity or severe dust.
Trouble Shooting and Error Messages
Limited Warranty
ADC
warrants its products against defects in materials and workmanship under normal household
use and service as follows:
1. Warranty service extends to the original retail purchaser only and commences with the date of delivery.
2. Your Adtemp Mini Thermometer is warranted for two years from date of purchase.
What Is Covered: Replacement of parts, and labor.
What Is Not Covered: Transportation charges to ADC. Damages caused by abuse, misuse, accident, or
negligence. Incidental, special, or consequential damages. Some states do not allow the exclusion or
limitation of incidental, special, or consequential damages, so this limitation may not apply to you.
To Obtain Warranty Service: Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Dr.,
Hauppauge, NY 11788. Please include your name and address, phone no., proof of purchase, and a brief
note explaining the problem.
Implied Warranty: Any implied warranty shall be limited in duration to the terms of this warranty and in
no case beyond the original selling price (except where prohibited by law). This warranty gives you specific
legal rights and you may have other rights which vary from state to state.
For Australian Consumers: Our goods come with guarantees that cannot be excluded under the Australian
Consumer Law. You are entitled to a replacement or refund for a major failure and compensation for any
other reasonably foreseeable loss or damage. You are also entitled to have the goods repaired or replaced
if the goods fail to be of acceptable quality and the failure does not amount to a major failure.
IB p/n 93-432-00 rev 2 (6/03/20) Printed in China
LATEX
K-JUMP HEALTH CO., LT D.
No. 56, Wu Kung 5th Road
New Taipei Industrial Park
New Taipei City 24890, Taiwan
POLYGREEN GERMANY GmbH
Ruhlsdorfer Straβe 95, D-14532,
Stahnsdorf, Brandenburg, Germany
Dist. by: ADC
®
55 Commerce Drive
Hauppauge, NY 11788
Inspected in the U.S.A.
Made in China
tel: 631-273-9600
toll free: 1-800-232-2670
fax: 631-273-9659
www.adctoday.com
0197
EC REP
Specifications
Measuring Site:
Forehead
Measuring Distance:
Within 0.7 in
(2 cm)
Measuring Time:
1 second
Measuring Range:
71.6°F to 109.9°F (22.0°C to 43.3°C)
Accuracy:
71.6°F to 109.9°F :0.5°F (22.0°C to 43.3°C :0.3°C )
Measuring Unit:
Dual (°F / °C)
Resolution:
0.1°F
Memory Sets:
9 sets
Operating Conditions:
60.8°F to 104°F (16°C to 40°C), 15 to 95% RH (non-condensing)
Storage Conditions:
-4°F to 122°F (-20°C to 50°C), 15 to 95% RH (non-condensing)
Battery:
Rechargeable Lithium Ion battery
Battery Life:
At least 300,000 measurements within warranty
Auto Shut Off:
After 10 seconds of inactivity
Dimensions:
3.48”x 1.02” x 1.02” (88.5mm x 26mm x 26mm)
Weight:
Approximately 1.59 oz (including battery)
Standards:
ISO 13485, IEC 62133, UN 38.3, ASTM E1965-98,
ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2
Read the instruction manual carefully before using this
device, especially the safety instructions, and keep the
instruction manual for future use.
Type BF applied part
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not
with domestic waste.
The reading is higher than measuring range, please repeat the measurement.
The reading is lower than measuring range, please repeat the measurement.
Out of operating condition.
Device self-diagnostic error, please contact customer service.
Low battery indication, please charge it via USB cable.
Temperature (°F) Beeps
T ≤ 98.9 2
99 ≤ T ≤ 99.4 2
T ≥ 99.5 8
Display Color
Green
Yellow
Red
Figure 1
USB for
Charging
93-432-00 rev 2.qxp_Layout 1 6/12/20 3:42 PM Page 2