4.1 GENERAL SPECIFICATIONS
Max Voltage between input terminal and Earth Ground: CATIII 600V.
Over-range Indication: display “1” for the significant digit.
Automatic display of negative polarity “–” .
Low Battery Indication: ‘ ’ displayed.
Max LCD display: 1999 (31/2 digits).
Fuse protection: F2-0.2A/250V (Ø5x20mm) F1-10A/500V (Ø6x30mm).
Power Supply: 9V battery, 6F22 or NEDA 1604.
Operating Temp.: 0°C to 40°C (relative humidity <85%).
Storage Temp.:-10°C to 50°C (relative humidity <85%).
Guaranteed precision Temp.: 23±5°C (relative humidity <85%).
Dimension: 150x74x43mm (covering the protective cover).
Weight: approx. 233g (including battery).
DC Voltage
Range Accuracy Resolution
200mV 0.1mV
2V ±(0.8% of rdg + 4 digits ) 1mV
20V 10mV
200V ±(0.8% of rdg + 5 digits) 100mV
600V ±(1.2% of rdg + 5 digits) 1V
AC Voltage
Range Accuracy Resolution
200V ±(1.2% of rdg + 20 digits) 100mV
600V ±(1.2% of rdg + 20 digits) 1V
Overload protection: DC 800V or 600V AC rms.
Input Impendance: 1MΩ.
Frequency Range: 40 to 400Hz.
Display: average value response (calibrated in rms of sine wave)
Input Impedance: 1MΩ.
Overload protection: 600V DC and AC 380V effective value.
8.6 TRANSISTOR TESTING
1. Set the measurement function range switch to the ‘hFE’ position.
2. Determine whether the transistor under testing is NPN or PNP. Insert the leads into the
correct holes of hFE socket on the front panel.
8.7 DIODE TEST
1. Connect the black test lead to COM probe socket and the red to VΩmA probe socket.
2.
Set the measurement function range switch to F position.
Note:
A) The meter will show approximate forward voltage drop of the diode.
8.8 CONTINUITY TESTING
1. Connect the black test lead to COM probe socket and the red to VΩmA probe socket.
2.
Set the measurement function range switch to the position.
Note:
If the circuit is open, figure ‘1’ will be displayed.
8. OPERATING INSTRUCTIONS
4. INTRODUCTION