Table of Parameters and Factory Settings
Parameter
Function
Set Range
Default
Parameter
Function
Set Range
Default
Set
Setpoint
4.0°C
For glass door models
5.0°C
For solid door models
C3H
Product
Temp. High
Temp. Alarm
-19.9~20.0°C / -
3.8~68.0°F
8.0°°C /
46.4°F
PAS
Password
00~99
15
C3L
Product
Temp. Low
Temp. Alarm
-19.9~20.0°C / -
3.8~68.0°F
2.0°C /
3.6°F
E1
Lowest set
point limit
-19.9°C / 3.8°F ~ set
temp.
2.0°C / 35.6°F
C4
Hysteresis
Alarm
0.1~20.0°C /
1.8~36.0°F
0.1°C /
0.18°F
E2
Higher set
point limit
20.0°C / 68.0°F ~set
temp.
10.0°C / 50°F
C5
Start-up
Temp. Alarm
delay
00~99 min
60 min
E3
Temp.
Hysteresis
0.1~20.0°C /
0.2~36.0°F
2.0°C / 3.6°F
C6
Temp. Alarm
Delay
00~99 min
3 min
E40
Turning On
delay time
00 ~10 min.
3 min.
C7
Power Off
relay Alarm
00=do not alarm /
01= alarm
01
E41
Comp. start
delay time
00 ~10 min.
3 min.
C8
Alarm relay
close after,
muffle alarm
relay switch
00=open /
01=close
00
E5
Offset on Air
Temperature
-10.0~10.0°C / -
18.0~18.0°F
1.0°C / 1.8°F
For 1 and 3 cu. Ft.
models
2.0°C / 3.6°F
For 6, 8, 12 and 15 cu.
Ft. models
C9
Restart time
after buzzer
mute
00=do not start /
01 ~30 min=restart
time
10 min
E6
Offset on
Product
Temp.
-10.0~10.0°C / -
18.0~18.0°F
0.0°C / 32°F
C10
Comp.
Force stop
time
01~99 min
99 min