Loading ...
Loading ...
Loading ...
Electromagnetic compatibility information
Electromagnetic compatibility information
40
41
Guidance and manufacture s declaration- electromagnetic emission'
The YE680B is intended for use in the electromagnetic environment
specified below. The customer or the user of the YE680B
should assure that it is used in such and environment.
Emission test
Compliance
Group 1
RF emissions
CISPR 11
Class B
RF emissions
CISPR 11
Class A
Complies
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Group1
For all ME EQUIPMENT and ME SYSTEMS
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Table 2
Guidance and manufacture s declaration-electromagnetic emission'
The is intended for use in the electromagnetic
environment specified below. The customer or the user of the
YE680B should assure that it is used in such and environment.
YE680B
Immunity
test
IEC 60601
test level
Compliance
level
Electrostatic
discharge
(ESD) IEC
61000-4-2
Electrical
fasttransient
/burst IEC
61000-4-4
±2 kV
100kHz repetition
frequency
Surge IEC
61000-4-5
±1 kV line(s)to line(s)
±2 kV line(s)to earth
±8 kV contact
±15 kV air
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 s
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
NOTE: UT is the a.c. mains voltage prior to application of the test level.
30A/m
50Hz or 60Hz
Power
frequency
(50/60 Hz)
magnetic
field IEC
61000-4-8
±8 kV contact
±15 kV air
±2 kV
100kHz repetition
frequency
±1 kV line(s)to line(s)
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 s
30A/m
50Hz or 60Hz
YE660 ( )
尺寸:180X142.5mm 色彩:准确、单色,层次分明
纸张:80克双胶纸 印后加工:对折装订
E CE认证 电子血压计说明书印刷要求
Loading ...