15
Operating
conditions
Temperature: 50°F-104°F (10°C-40°C)
Humidity: 15-95% RH, non-condensing
Atmospheric pressure: 860hPa-1060hPa
Storage and
Shipping
Conditions
Temperature: -F-122°F (-20°C-50°C)
Humidity: 0-95% RH, non-condensing
Atmospheric pressure: 860hPa-1060hPa
Type of measuring
Applicable for forehead temperature and ear
temperature
Battery Changeable for two 1.5V triple AAA batteries
Battery life More than 3000 times
Accuracy for clinical
test
The maximum allowable error for clinical test is
specified in the formula below:
0.F/0.C (for 95% )
Error! Bookmark not defined.Error! Bookmark
not defined.
Among which:
T1 and T2 represent temperature value for
thermometer under test respectively, Tref
represents the constant reference temperature
This infrared thermometer meets requirements established in ASTM
Standard (E 1965-98). Full responsibility for the conformance of this
product to the standard is assumed by Innovo Medical, 406 Sterling
Street, Building 2, Camphill, PA 17011, USA.
ASTM laboratory accuracy requirements in the display range of 98°F
to102°F (37°C-39°C) for IR thermometers is ±0.4°F (±0.2°C), whereas for
mercury-in-glass and electronic thermometers, the requirement per
ASTM Standards E 667-86 and E 1112-86 is ±0.2°F (±0.1°C).
Terror =
T1-Tref
T2-Tref
+
2