Measurement Range 32.0ºC~42.9ºC (89.6~109.2ºF)
Display LCD 3 1/2 digits with ºC
Display resolution 0.1ºC or 0.2ºF
Accuracy ±0.1°C 35.5°C~42.0°C(±0.2°F 95.9°F~107.6°F)
±0.2°C under 35.5°C or over 42.0°C
(±0.4°F under 95.9°F or over 107.6°F)
at standard room temperature 25.0°C (77.0°F)
Memory For storing the last measured value
Battery One 1.5 V DC. button size battery (SR41 or
LR41)
Battery Life Approx. 1500 times operation or 1 year with
1-2 measurement per day including standby
mode.
Dimension TG250 : 13.2x2.3x1.3cm
TG380 : 13.2x2.3x1.3cm
Weight TG250 : 14.5 grams including battery
TG380 : 14.5 grams including battery
Alarm beep sounds(beep-beep-beep 10 times) when
peak temperature is reached
Environmental for
using
Temperature: 5~40ºC (41ºF~104ºF)
Humidity: ≤95%RH
Waterproof Level Waterproof
Storage and trans-
portation condition
Temperature: -25.0°C~55.0°C (-13.0°F~131.0°F)
Humidity: ≤95%RH
Accessories battery, carry case, instruction manual
Safety classication Type B Applied Part
Guidance and manufacturer’s declaration-electromagnetic immunity
The TG250/TG380 is intended for use in the electromagnetic environment specied below.
The customer or the user of the TG250/TG380 should assure that is used in such and environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2,5 GHz
Not applicable
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the TG250/
TG380 including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from xed RF transmitters, as deter-
mined by an electromagnetic site survey.
a. Should be less than the compliance level in each
frequency range.
b. Interference may occur in the vicinity of equipment
marked with the following symbol:
Specications
EMC guidance and manufacturer’s declaration
This instrument is covered by a 5 year guarantee from the date of
purchase, batteries and accessories are not included. The guaran-
tee is valid only on presentation of the guarantee card completed
by the dealer conrming date of purchase or the receipt. Opening
or altering the instrument invalidates the guarantee. The guaran-
tee does not cover damage, accidents or non-compliance with the
instruction manual. Please contact rossmax service.
Warranty
IN0TG2500000000XX
OBM_IB_TG250_TG380_EN_ver1303
Product Information
Date of purchase:
Store where purchased:
Price Paid (excl. Tax):
Purchase for:
Guidance and manufacturer’s declaration-electromagnetic emissions
The TG250/TG380 is intended for use in the electromagnetic environment specied below.
The customer or the user of the TG250/TG380 should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions CISPR 11 Group 1 The TG250/TG380 uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equip-
ment.
RF emissions CISPR 11 Class B The TG250/TG380 is suitable for use in all establishments,
including domestic establishments and those directly con-
nected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage uctuations/
icker emissions IEC
61000-3-3
Not applicable
Guidance and manufacturer’s declaration-electromagnetic immunity
The TG250/TG380 is intended for use in the electromagnetic environment specied below.
The customer or the user of the TG250/TG380 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%
Electrical fast
transient/burst
IEC61000-4-4
+ 2kV for power
supply lines
+ 1kV for input/
output lines
Not applicable
Not applicable
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Surge IEC 61000-
4-5
+ 1kV line(s) to
line(s)
+ 2kV line(s) to
earth
Not applicable
Not applicable
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
<5% UT(>95% dip
in UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip
in UT) for 5 s
Not applicable
Not applicable
Not applicable
Not applicable
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the TG250/TG380
requires continued operation during power
mains interruptions, it is recommended
that the TG250/TG380 be powered from
an uninterruptible power supply or a bat-
tery.
Power fre-
quency(50/60 Hz)
magnetic eld IEC
61000-4-8
3 A/m 3 A/m The TG250/TG380 power frequency
magnetic elds should be at levels char-
acteristic of a typical location in a typical
commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Recommended separation distance between portable and mobile RF communications equipment and
the TG250/TG380
The TG250/TG380 is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of the TG250/TG380 can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the TG250/TG380 as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d =1,2√P
80 MHz to 800 MHz
d =1,2√P
800 MHz to 2,5 GHz
d =2,3√P
0,01 N/A 0,12 0,23
0,1 N/A 0,38 0,73
1 N/A 1,2 2,3
10 N/A 3,8 7,3
100 N/A 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the trans-
mitter, where P
The maximum output power rating of the transmitter in watts (W) according to the transmitter manu-
facturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reection from structures, objects and people.
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reection from structures, objects and people.
a Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios,amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre-
dicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmit-
ters, an electromagnetic site survey should be considered. If the measured eld strength in the location
in which the TG250/TG380 is used exceeds the applicable RF compliance level above, the TG250/
TG380 should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the TG250/TG380.
b Over the frequency range 150 kHz to 80 MHz, eld strengths should be les than 3 V/m.
www.rossmax.com
Rossmax International Ltd.
12F., No. 189, Kang Chien Rd., Taipei, 114,
Taiwan.
Rossmax Swiss GmbH,